Irradiador Gamma con Soporte de Calibración UDG-AT110

UDG-AT110AT110 Gamma beam irradiator with calibration bench is designed for verification, calibration and examination of ionization radiation measurement instruments in collimated field of gamma radiation.

 

 

 

 

 

 

Descripción

Operating principle

Irradiator based on the use of radionuclide sources of 137Cs.

The installation scheme is realized with a fixed irradiator and linear-positioned platform of calibration bench.
The range values ??of dose rate of gamma radiation is achieved by using different sources of 137Cs with different activity and changing in the distance "a source-detector".

Sizeof the radiation field varies by distance "a source-detector" or diameter collimator channel.
The positioning of calibrated equipment in the collimated gamma beamis provided by the calibration stand.
Management function of the stand is carried out by means of software and portable control panel.

UDG-AT110.JPG2

Application

  • Metrological assurance of gamma radiation dosimetry:
    • Laboratories for research, adjustment and serial production of dosimetric instruments
    • Metrology laboratories focused on calibration and verification of dosimetric instruments
    • Secondary standard dosimetric laboratories (SSDL)


Features

  • Up to 5 gamma sources in irradiator protection container
  • Software, motion control sources in theirradiator
  • Automatic and manual positioning
  • Movingsources in the irradiator and the moving platformusing digital servos
  • Centering the reference and calibrated equipment using laser systems
  • Remote monitoring the indicated value of calibrated instruments using video surveillance system
  • Safe braking and travel limitation of movable platform
  • Lock system and visual alarms of source position
  • Continuous control of radiation environment
  • Emergency power sources
  • Download the sources in the installation using transferring container and accessory set

Especificaciones

Used gamma radiation sources, maximum activity 137Cs – 1.3·1012 Bq (35 Ci)
Range of formed kerma rates 0.25 μGy/h – 350 mGy/h
Collimation unit ø60 mm or 90 mm, length 150 mm
Elevation of radiation beam axis (1500±30) mm
Accuracy of positioningthe sourceheight ± 1 mm
Time to transfersourceto the operating position no more 15 s
Own radiation background in the mode “storage”, 0.50 μSv/h
at a distance 1 m, does not exceed  
Basic error ± 4 - 7 %
Operating distance range(R) 0.5 - 8 m
Reproducibilityofthe movingplatform by coordinate X less than 0.5 mm
Display stepof working distance 0.1 mm
Absolute error of positioning not more than 0.002 R
Platform travel speed from 0.9 mm/s to 26 cm/s
The range of operational movementsof moving table:  
          vertically from the floor from 1250 to 1550 mm
          horizontal:  
                    along the radiation beam axis horizontal (axis Y) ± 50 mm
                    across the radiation beam axis (axis Z) ± 140 mm
          around a vertical axis in increments of 15º 360º
Weightof equipment installed:  
          on moving table up to 35 kg
          on moving platform up to 75 kg
Operation mode setup time not more than 1 min
Continuous operation time not less than 24 h
Power supply       230 (±23) V, (50 ± 1) Hz
Required power        not more than 600 V·A
Operating temperature range from +15°C to +25°C
Relative humidity up to80%
Dimensions, no more:  
          irradiator 640x640x1700 mm
          base 8000x860x200 mm
          movable platform 910x855x1820 mm
          movable table 270x330 mm
          operator's workplace (sq.) 3500x1500 mm
Weight, no more:  
          irradiator 735 kg
          base 135 kg
          movable platform 70 kg
          transferring container 100 kg
          operator's workplace 150 kg

 